Dosimetric comparison between helical tomotherapy and VMAT for pediatric total lymphoid irradiation
Carlos Ferrer, Spain
PO-1733

Abstract

Dosimetric comparison between helical tomotherapy and VMAT for pediatric total lymphoid irradiation
Authors:

Carlos Ferrer1, Concepción Huertas1, Abrahams Ocanto2, David García3, Anne Escribano2, Moisés Sáez1

1HU La Paz, Medical Physics, Madrid, Spain; 2HU La Paz, Radiation Oncology, Madrid, Spain; 3HU La Paz, Medicla Physics, Madrid, Spain

Show Affiliations
Purpose or Objective

The aim of this study is to compare the planning of volumetric modulated arc therapy (VMAT) treatments with those planned with helical tomotherapy (HT) in pediatric total lymphoid irradiation (TLI).

Material and Methods

TLI treatments are used in the management of pediatric allogeneic hematopoietic stem cell transplantation (HSCT). TLI treatment was prescribed to 8 Gy in 4 fractions, and planned on an Elekta Infinity accelerator with a 160-leaf Agility collimator and VMAT technique with Elekta Monaco TPS (v. 5.11.02). The same treatment was planned on a HD helical Tomotherapy, and Accuray Precision TPS. Both linacs have 6 MV nominal energy. For organs at risk (OAR), in the treatment planned with Monaco, a single cost function named conformality was used to conform the dose, whereas in the treatment planned with tomotherapy, multiple auxiliary volumes were created and several cost functions were used for each one of them, which makes this planning more complex. The dose objective is that 100% of the PTV is covered by at least 95% of the prescribed dose. Conformality (CI) and homogeneity (HI) indices are compared between HT and VMAT, as well as doses to OAR.

Results

The PTV coverage requirement was met for both treatment planning systems (V100% ≥ 95%), being slightly higher for VMAT, 96.6% (VMAT) vs 95% (HT). The treatment planned in HT presents better CI (1.05 (HT) vs 1.34 (VMAT)) and the dose slightly more homogeneous (HI= 0.09 (TH) vs 0.14 (VMAT)). The doses to OARs, shown in the table (in cGy), were similar for both planning systems. Treatment time was significantly longer in HT, 20.2 min (TH) vs 13.7 min (VMAT).

Conclusion

Planning with HT showed better dose distribution conformity and homogeneity compared to planning with VMAT. HT planning was superior in terms of reduction of mean and maximum doses in practically all the OAR. However, the differences between both are small, with the mean dose in some OARs slightly lower with VMAT, such as in the left kidney. Longer treatment time in HT could increase uncertainty due to intrafraction motion.