In-silico assessment of the out-of-field over-response of an optically stimulated luminescent film
Marco Caprioli, Belgium
PO-1539

Abstract

In-silico assessment of the out-of-field over-response of an optically stimulated luminescent film
Authors:

Marco Caprioli1, Marijke De Saint-Hubert2, Luana de Freitas Nascimento3, Laurence Delombaerde4, Shamil Samanta Galvez Febles5, Katleen Himschoot6, Dirk Vandenbroucke6, Paul Leblans6, Wouter Crijns7

1KU Leuven, Oncology, Leuven, Belgium; 2Belgian Nuclear Research Centre (SCK CEN), Research in Dosimetric Application, Mol, Belgium; 3Belgian Nuclear Research Centre (SCK CEN), Research in Dosimetric Application , Mol, Belgium; 4KU Leuven , Oncology, Leuven, Belgium; 5Università degli Studi di Milano, Fisica, Milano, Italy; 6Agfa N.V., Corporate Innovation Office, Mortsel, Belgium; 7University Hospital Leuven, Department of Radiation Oncology, Leuven, Belgium

Show Affiliations
Purpose or Objective

The QUARTEL project aims to optimize 2D Optically Stimulated Luminescent (OSL) film for pre-treatment radiotherapy dosimetry. This dosimeter has an energy dependent signal resulting in a varying response in function of position (in-field vs out-field), beam-modulation (static beams vs IMRT), etc. This work investigates in-silico evaluation of the out-of-field over-response to avoid measurements-based assessment using existing analytical Z-effective calculations (ZeffC) and Monte Carlo (MC) dose simulations. 

Material and Methods

In the context of the QUARTEL project three OSL films were produced utilizing a BaFBr:Eu2+ powder embedded in a water equivalent polymer-based binder. Three samples with different weight concentration were tested:  5,2% (A), 6.5% (B), 10% (C). For the in-silico assessment, these phosphor films were first modeled as homogeneous distributions of BaFBr:Eu2+. The OSL over-response assessment uses (1) a complete out of field energy spectrum (setup1: Varian 6MV, 10cm off-axis, 10x10 cm2 field, 0° gantry angle, 90 cm SSD, 5cm depth) and (2) two mono-energetic spectra representing the in-field (1.25MeV) and out-of-field (0.115 MeV) situations for both ZeffC and MC simulations (MCNP 6.2.)The over-response in setup 1) was measured in a Varian Linac and considers the mean value of a 1x1 cm2 ROI at 7cm from the central axis due to the limited size of the film. All the data (Zeff, dose) were divided by the corresponding response in water as in (1) and (2).  

Results

The measured over-response is consistent with the in-silico assessment (Figure 1 and 2), although the difference between samples A and B is comparable with the standard deviation within the ROI: 0.0069. The calculated Zeff is 1.48 (A), 1.53 (B), and 1.77 (C) times higher than water. Similarly, for MC, the simulated dose response increases with the phosphor (%) from 1.77 (A) to 2.45 (C) with a precision of 0.29%. The in-silico results passing from the represented in-field to the out-of-field regions (2) show an increased over-response at lower energies (Table 1) up to 2.41 times for the MC simulated sample C. 


Table 1. In-silico assessment of the OSL out-of-field over-response (R) in 2 monoenergetic beams representative for an out-of-field and in-field photon energy spectrum.

Phosphor concentration (%)ROSL/RWater(115 KeV)ROSL/RWater (1.25 MeV)
5.2 (A)1.61(MC)/1.54(Zeff)0.91(MC)/1.15(Zeff)
6.5 (B)1.84(MC)/1.61(Zeff)
0.91(MC)/1.13(Zeff)
10 (C)2.17(MC)/2.00(Zeff)
0.90(MC)/1.20(Zeff)

 

Conclusion

The in-silico assessment of the out-of-field over-response shows a correlation with the Z-effective calculation as well as the MC simulations. Both methodologies can be utilized to evaluate the energy dependence for new OSL material. The Z-effective could be used for a quick assessment, but MC can cope with non-homogeneous layers as well (future work). 

The out-of-field spectra were provided by the team of Prof. Dr. Stephen Kry of MD Anderson.

The Zeff calculator is produced by ML Taylor, et al., Med. Phys. 39 (2012) 1769-1778